Leica Cryo-CLEM Imaging Protocol
Published:
1. Pre-Experiment Preparation
a. Transfer Shuttle Drying
- Dry the Transfer Shuttle on the Leica Cryo Tool Dryer (CTD) for at least 20 minutes before use to reduce contamination.
b. Cryo Stage Heating
- Run stage heating for 30 minutes at 60°C:
- Do not connect the LN₂ hose.
- Lift the objective to the uppermost position (fully retracted).
2. Power On Sequence
a. Microscope
- Turn on the microscope, the control panel and the pump.
- Turn on the camera.
b. Laser
- Important: Turn on the front switch of the laser before the back switch.
- Otherwise, the laser will not power on.
3. Launch Imaging Software
- Start LAS X software.
4. Initialize Motorized Stage
- Click Initialize Stage.
- ⚠️ Do not block the stage — allow it to move freely during initialization.
5. Select Imaging Mode
- Navigate to:
MatrixScreener → Matrix MAPS + CLEM
6. Fill Liquid Nitrogen (LN₂)
- Fill LN₂ into the storage tank.
- ⚠️ Do not fill all the way to the top:
- LN₂ may spill when inserting the heat exchanger.
7. Connect Cryo Hose to Pump and Stage
a. Start the Pump
- Press COOL to start.
b. Wait for Purge
- Wait until the hose is purging nitrogen gas.
c. Connect Hose to Cryo Stage
- Insert the hose into the cryo stage.
- Lock the hose in place:
- Push the locking screw toward the hose while tightening.
- ⚠️ If not pushed first, the hose may be pulled out easily and won’t seal properly.
- Moisture may enter and condense, blocking the view.
8. Wait for Stage Cool Down
- When temperature reaches -195°C, wait another 15 minutes before loading the sample.
9. Cool Down the Transfer Shuttle
a. Setup
- Ensure the transfer shuttle shutter is closed.
b. Insert Cartridge
- Place an empty cartridge into the loading position.
- Close the slider.
- Turn the rod to the open-O position.
- Cover the cartridge with the shielding.
c. Fill LN₂
- Cover the transfer shuttle with its lid.
- Open the rotatable opening.
- Fill LN₂ inside the shuttle.
- ⚠️ Do not overfill — avoid pouring over the rod.
- Otherwise, the rod may freeze and become immobile.
- The shuttle is fully cooled when the LN₂ stops boiling inside.
10. Load AutoGrids into Cartridge
- Proceed with loading autogrids into the pre-cooled cartridge.
11. Load Grid into Microscope
a. Ensure Lens Is Up
- ⚠️ Objective lens must be lifted to avoid damage from the shuttle.
b. Stage Access
- Open the stage slider.
c. Insert Transfer Rod
- Insert the transfer rod fully.
- Rotate to Open position to release the cartridge.
- ⚠️ Must insert all the way to the end.
- Otherwise, you won’t be able to focus on the grid.
d. Retract Rod
- Pull the rod back.
- The cartridge should remain in the stage.
e. Clean Up Shuttle
- Remove the transfer shuttle from the stage.
- Pour out any remaining LN₂.
- Place it on the Cryo Tool Dryer (CTD) for baking out.
12. Imaging Procedure
a. Load Experiment
- Open LAS X and load your experiment.
b. Move to Left Grid
- In the MatrixScreener window:
- Right-click the left grid.
- Select “Move stage to left grid center”.
c. Autofocus
- Run Auto Focus.
- ⚠️ If autofocus fails, adjust manually until sharp.
d. Spiral Scan
- Run the Spiral Scan to locate good regions for imaging.
e. Focus Map Setup
- Select a circular imaging area.
- Click on several points within the circle to define a focus map.
- Click Start Focus Map to collect Z positions.
f. Mosaic Scan
- Enable 4 channels.
- Adjust the Z-stack range according to your sample thickness.
- Click Run Matrix to begin automated imaging.
g. Export Data
- After acquisition is finished, export data files for downstream analysis.
13. Finishing Up
a. Close Software
- Exit the LAS X software and any other imaging applications.
b. Retrieve Grids
- Lift the objective lens to its uppermost position.
- Retrieve the cryo grids from the stage.
c. Heat the Stage
- On the temperature control box, press “Heat”.
- Wait until the stage heats to 60°C.
⚠️ Do not remove the hose before the temperature reaches 60°C.
Removing the hose prematurely can cause damage due to freezing.
d. Disconnect the Cryo System
- Once the stage is at 60°C:
- Remove the LN₂ hose.
- Remove the pump from the LN₂ tank.
e. Return Grid Boxes
- Place the used grid boxes back into the LN₂ storage tank for long-term storage.

